Gauge for oil refractometer
diopter oil is used to calibrate the refractometer with whipped performance scale. Packed in glass bottles of 1 ml.
Set
diopter oil – 1.
Calibration unit – 1.
How to calibrate refractometer dioptrytychnym butter and calibration unit?
The glossy surface gauge prism diopter drip 1-2 drops of oil. Wait to rozteklos oil evenly. Warning prism surface completely be covered with oil gauge (without spaces or air bubbles).
Calibration unit with oil lay on the glossy side surface of the refractometer measurement.
Light pressure unit to push refractometer prisms. Within 30 seconds the temperature leveled and block device.
Turn calibration screw to adjust the scale refractometer (for Brix 78.4)
refractometer calibration is performed often. Therefore diopter bottle of oil will last for decades.
Features
Oil Volume (ml): 1
Block size: 1,4h0,8h0,5